Trillium 120PA
Image

The new Trillium 120PA version is identical in form, function and performance to the T120P and now is available with the added benefit of fast "one touch" motorized mass centering. Motorized mass centering simplifies seismometer installation and gives customers the option of remotely re-centering the masses via a telemetry connection.

This seismometer is an exceptional instrument with a self-noise within 4db of the NLNM at 100 seconds and below the NLNM up to 10 Hz.

This instrument incorporates the same symmetrical triaxial design and suspension system as the highly successful Trillium 240. The robustness and reliability of the mechancial suspension is well-proven; with over 1000 Trillium units operating in the field, there have been no suspension system failures.

As with all Trillium sensors, the 120PA does not have a mass lock. Instead the T120PA uses a field proven suspension system capable of withstanding the rigors of field handling and transportation. During deployment the sensor requires only manual leveling before operation.

Trillium 120PA Technical Specifications

 TECHNOLOGY                                                               
 Topology Symmetric triaxial
 Feedback Force balance with capacitive transducer
 Mass centering
Automatic motorized re-centering, can be remotely initiated 
 LevelingIntegrated bubble level; Adjustable locking leveling feet 
 Alignment
Vertical scribe marks for (N and S); Precision guide in cover for straight-edge,
line or laser level; Holes for 5/16" alignment (E and W)
  
 PERFORMANCE 
 Self-noise Below the NLNM 35 s up to 10 Hz
 Sensitivity1200 V-s/m ±0.5% precision, Contact factory for other options
 Bandwidth-3 dB points at 120s and 145 Hz 
 Clip level> 15 mm/s up to 1.5 Hz 
 Temperature±45°C without re-centering
  
 INTERFACE 
 Connector19-pin MIL-C-28642
 Velocity output40 V peak-to-peak differential; Selectable XYZ or UVW mode 
 Mass positionThree independent voltage outputs 
 Calibration input
Single voltage input with one active high control signal per channel; Calibration
in XYZ or UVW
 Serial port

For enhanced instrument control and status:

  • Mass center, UVW/XYZ mode, short/long period mode, firmware updates
  • Temperature, mass position, instrument status, serial number 
  
 POWER 
 Supply voltage9 to 36 volts DC isolated input 
 Power consumption620 mW typical at 15 V input 
 Protection
Reverse-voltage protection; Auto-resettable over-current protection
(No fuse to replace)
  
 PHYSICAL 
 Diameter21.0 cm 
 Height21.4 cm ±0.5 cm depending on leveling feet extension 
 Weight7.5 Kg 
 HandlingDetachable carrying handle on lid
  
 ENVIRONMENTAL 
 Operating temp.-20°C to +50°C
 Humidity0 to 100% 
 Shock
20g half sine, 5 ms without damage, 6 axis;
No mass lock required for transport 
 PackagingRated to IP68 and NEMA6P for outdoor use 